Applications
GenX replacement PFOA compound by LC-MS/MS
25436
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Environmental
Product / Trade Good
Chemical Compound
Chemical Compound
Environmental / Health

Applications

GenX replacement PFOA compound by LC-MS/MS
Analytes
1
2
3
4
5
6
7
8
9
Compound Name
Cid429733
Details
LC Conditions (App ID: 25436)
Column:
Brand Name:
Part No:
Elution Type:
Gradient
Mobile Phase:
A: 20mM Ammonium Acetate in Water
B: Methanol
Gradient Profile:
Step No.Time(min)%A%B
10955
20.14555
34.5199
48199
58.5955
Flow Rate:
0.6 mL/min
Temperature:
40°C
System:
Shimadzu Shimadzu
Detection:
LC/MS/MS @ 0°C
Sample Preparation Details
Description:
Strata GCB, 500mg/6mL, 30/Pk
Part Number:
Pre Treatment Note:
All standards including HFPO-DA and its stable isotope labelled surrogate 13C3-HFPO-DA were purchased from Wellington Laboratories (Guelph, Ontario). 200 mL water samples were extracted using Strata®-XL-AW weak anion exchange SPE cartridges following the conditions in the ISO standard 25101. Samples were also passed through Strata GCB following the requirements of DOD QSM 5.1. The final volume of the SPE eluent was 8 mL. Final extracts consisted of 80% methanol and 20% water. Solid Phase Extraction Protocol Cartridge: Strata XL-AW 500 mg/6 mL Part No.: (8B-S051-HCH) Condition: 6 mL 0.3% Ammonium hydroxide in Methanol, followed by 6 mL Methanol Equilibrate: 6mL Water Load: Add sample at about 10 mL/min Note: use an adapter cap (Part no.: AH0-7191) and large volume sample reservoir (Part No.: AH0-7005) Wash: 2x 6mL Water Elute: 10 mL 0.3% Ammonium hydroxide in Methanol(this 10 mL will elute PFAS/GenX analytes) Load: SPE Eluent through Strata GCB 250 mg/6 mL (8B-S528-HCH) Evvaporate: to dryness and reconstitute to 1.0 mL with Methanol/ Water (96:4)(containing internal standards)